Build-Up Detection
Build-Up Index
Select mode for build-up index.
The interval for determining the build-up measured value can be specified here.
- If experience shows that build-up forms quickly, then a fast interval should be selected. Also for cleaning cycles.
- If build-up forms over months, then a slow or standard interval is sufficient.
Measurement every:
- Slow: 600 sec.
- Standard: 60 sec.
- Fast: 6 sec
TIP:
The build-up index is only available in conjunction
with Heartbeat Technology. If Heartbeat Technology was ordered together with the measuring
device, the option will already be enabled, and no further action is required. If Heartbeat
Technology was ordered at a later date, you must first activate the option under "Activate
SW option" by entering the activation key you received. To purchase Heartbeat Technology,
contact your local sales and service center. In addition to Heartbeat Technology,
conductivity measurement must be enabled on the device. To do this, go to the "Conductivity
measurement" parameter on the "Process parameters" menu and select the "On" option.
Property | Value |
|---|---|
Selection / Range |
|
Factory Setting | Off |
Visibility | Operator |
Write Access | Maintenance |
Data Type | ENUM16 |
Build-Up Index Damping
Enter damping value for build-up index
TIP:
The build-up index is only available in conjunction
with Heartbeat Technology. If Heartbeat Technology was ordered together with the measuring
device, the option will already be enabled, and no further action is required. If Heartbeat
Technology was ordered at a later date, you must first activate the option under "Activate
SW option" by entering the activation key you received. To purchase Heartbeat Technology,
contact your local sales and service center. In addition to Heartbeat Technology,
conductivity measurement must be enabled on the device. To do this, go to the "Conductivity
measurement" parameter on the "Process parameters" menu and select the "On" option.
Property | Value |
|---|---|
Selection / Range | 0 - 15 |
Factory Setting | 0 |
Visibility | Operator |
Write Access | Operator |
Data Type | UINT8 |
Build-Up Index Value
Shows current build-up index value
TIP:
The build-up index is only available in conjunction
with Heartbeat Technology. If Heartbeat Technology was ordered together with the measuring
device, the option will already be enabled, and no further action is required. If Heartbeat
Technology was ordered at a later date, you must first activate the option under "Activate
SW option" by entering the activation key you received. To purchase Heartbeat Technology,
contact your local sales and service center. In addition to Heartbeat Technology,
conductivity measurement must be enabled on the device. To do this, go to the "Conductivity
measurement" parameter on the "Process parameters" menu and select the "On" option.
Property | Value |
|---|---|
Selection / Range | 0.0 - 3.4E+38 % |
Factory Setting | 0.0 % |
Visibility | Operator |
Write Access | Read-Only |
Data Type | FLOAT |
Build-Up Limit
Enter limit value for the build-up index
TIP:
The build-up index is only available in conjunction
with Heartbeat Technology. If Heartbeat Technology was ordered together with the measuring
device, the option will already be enabled, and no further action is required. If Heartbeat
Technology was ordered at a later date, you must first activate the option under "Activate
SW option" by entering the activation key you received. To purchase Heartbeat Technology,
contact your local sales and service center. In addition to Heartbeat Technology,
conductivity measurement must be enabled on the device. To do this, go to the "Conductivity
measurement" parameter on the "Process parameters" menu and select the "On" option.
Property | Value |
|---|---|
Selection / Range | 0 - 100 % |
Factory Setting | 50 % |
Visibility | Operator |
Write Access | Operator |
Data Type | FLOAT |
Build-Up Limit Hysteresis
Enter hysteresis for build-up limit value
TIP:
The build-up index is only available in conjunction
with Heartbeat Technology. If Heartbeat Technology was ordered together with the measuring
device, the option will already be enabled, and no further action is required. If Heartbeat
Technology was ordered at a later date, you must first activate the option under "Activate
SW option" by entering the activation key you received. To purchase Heartbeat Technology,
contact your local sales and service center. In addition to Heartbeat Technology,
conductivity measurement must be enabled on the device. To do this, go to the "Conductivity
measurement" parameter on the "Process parameters" menu and select the "On" option.
Property | Value |
|---|---|
Selection / Range | 0 - 100 % |
Factory Setting | 20 % |
Visibility | Operator |
Write Access | Operator |
Data Type | FLOAT |
Provide Feedback