Diagnostic Events
The Process Events can be set in the Event Configuration.
By default no event is activated, so the message “Event disabled in configuration” is shown.
If it is configured but no event is active, the message “No event active” is shown. In case of
an event, the set message is displayed.
Special event in the process | Meaning | Special condition |
|---|---|---|
Not totally filled | The tube is not totally filled. If the parameter Refresh time is set to Very short, the event Not totally filled
cannot be monitored. | Not all the sensors are in contact with the liquid. |
Liquid out of range | The speed of sound in the liquid is out of range. |
|
Unstable flow | The flow rate is not stable. | The standard deviation of the flow rate measurements is too high. |
Low flow cut off | The cut-off value of the flow rate has been used. | The cut-off function must be enabled |
Change of liquid | A different liquid flows in the pipe. The message is active for 10 s on the
display. | The speed of sound in the liquid has changed by more than 3 m/s in 1 second. |
Backward flow | The liquid flows in the opposite direction as the one set as positive direction
of the flow. |