Volume & Mass Totalizers

Counting Direction

By default, the counting direction of all totalizers is Positive Only.
The possible counting directions are:
Totalizer counting direction selections
Counting direction
Totalizer function
Positive Only
The totalizer counts the quantity of liquid that flows in the direction defined as positive, i.e. in the same direction as the arrow located on the front of the device.
Negative Only
The totalizer counts the quantity of liquid that flows in the direction defined as negative, i.e. in the direction opposite to the direction of the arrow located on the front of the device.
Positive and Negative
The totalizer counts the quantity of liquid that flows in the direction defined as positive but deducts the quantity of liquid that flows in the direction defined as negative